- Product Details
Mooney Viscometer viscosity for rubber vulcanization index determination. Sample in a certain temperature and pressure, the rotor to a certain degree of rotation torque on the sample to a certain degree of shear stress, measure out the rubber anti-shear torque. Is reclaimed rubber, rubber, wire and cable industry indispensable instrument.
Rubber Mooney Viscometer Design Principle
Place the test rubber with a specified pressure and maintain a set temperature of the mold cavity, so that the rotor speed in the cavity filled with rubber 2r/min rotation. Determination of the friction of rubber, when the rotor rotates, and this torque required to convert the data values in Mooney value calculation unit. Mooney scorch test, the test rubber viscosity changes, determination of the plastic material of the test to reach a certain curing time and curing rate of the extent required. 1.With the start of the trial, the viscosity of the rubber sample size changes;All the test process can be real-time display of the screen, as well as see the curve corresponding to a Mooney viscosity and time(Mooney viscosity curve or the Mooney scorch curve),curve shape and characteristics of the rubber material and test temperature.After the test, the test data, test curve will be automatically stored in the database.For printing, just click on the print . 2.Obtain the following data values from the test report: a.initial value b.Mooney value【M(P+N):P--Warm-up time(min) N—Test time(min)】; c.T5,T35,Δt30(small rotor:T3,T18, Δt15); d.The actual temperature curve of the upper and lower cavity e.Test date and time; f.Other: the value of relaxation, the number of plastic material, the operator (input).
Meet the standards:
ASTM D1646
6. ISO 289-1 (1994) Unvulcanized rubber - Determination with a shear disc viscometer
Applicable industry:
HZ-7001B Mooney Viscometer widely used in the measurement of quality inspection; rubber research; automotive manufacturing; wire and cable; medical equipment; civil aviation; institutions of higher learning; research laboratories; technical supervision departments; petrochemical; tire production; other industries.
Technical parameter:
Model | HZ-7001B Mooney Viscometer |
Control temperature range | Normal temperature~200°C |
Temperature display resolution | 0.01°C (0.1°C common type of imported temperature controller temperature) |
Temperature control accuracy | ±0.3°C |
Calibration accuracy | Mooney value ±0.5 |
Rotor speed | 2 rotation/ min ±0.02rotation/ min |
Torque range | Mooney value 0~200 |
Torque resolution | Mooney value 0.1 |
Print content | Date, time, temperature, Mooney scorch curve and |
Power | 50HZ, ~220V±10% |