- Product Details
The walk-in constant temperature and humidity test room is mainly composed of a control panel, a switchboard, a moisturizing panel blower, a heater, a humidifier, and a freezer. It provides the industry with a large temperature and humidity test environment for large parts, semi-finished products, and finished products. It is suitable for testing large-volume test equipment such as computer terminals and automotive parts.
Comply with the standards
1. GB11158 High temperature test chamber technical conditions
2. GB10589-89 Low temperature test chamber technical conditions
3. GB/T10586-89 Humidity test chamber technical conditions
4. GB/T2423.1-2001 Low temperature test chamber test method
5. GB/T2423.2-2001 High temperature test chamber test method
6. GB/T2423.3-93 Humidity test chamber test method
7. GB/T2423.4-93 Alternating humidity test method
8. IEC60068-2-1.1990 Low temperature test chamber test method
9. IEC60068-2-2.1974 High temperature test chamber test method
10. GJB150.3 High temperature test
11. GJB150.4 Low temperature test
12. GJB150.9 Damp heat test
13. GB/T2423.22 Temperature change test
Technical Parameters:
Model | HZ-Volume (Low Temperature Selection: A~D) | |
Size (cm) | Designed according to the size of the space specified by the customer | |
Temperature range | -65℃~+80℃ | |
Humidity range | 20%~95%R.H. | |
Temperature analysis accuracy/uniformity | ±0.1℃;±0.1%R.H./±1.5℃;±5.0%R.H. | |
emperature control accuracy/fluctuation | ±1.0℃;±2.0%R.H./±0.5℃;±3.0%R.H. | |
Cooling time | About 4.0℃/minute; about 0.7℃~1.0℃/minute | |
Internal and external materials | Outer box is painted steel plate or stainless steel plate with matte finish, inner box is stainless steel | |
Insulation material | High temperature resistant high density ethyl chloroformate foam insulation material | |
Cooling system | Water-cooled/two-stage compressor |