Troubleshooting: How to deal with water leakage in constant temperature and humidity testing chamber?

Troubleshooting: How to deal with water leakage in constant temperature and humidity testing chamber?
If the long-term use of constant temperature and humidity test chamber does not pay attention to maintenance and maintenance, it is prone to various large or small failures. The constant temperature and humidity test box in some cases will present the situation will be different, however, in the test process some failure problems will also occur frequently, such as water leakage, so, when the constant temperature and humidity test box water leakage how to do? Here Xiaobian Gao Tian explains the reasons and treatment methods.

1, you need to check the water of the constant temperature and humidity test box, see if the water valve is opened too large, resulting in sufficient water flow, if so, the water valve can be turned down.

2, check whether the constant temperature and humidity test chamber water supply time is set too often, resulting in too much water, if this situation, you need to expand the water supply interval, so that the frequency of water will be reduced.

3, check the constant temperature and humidity test box place is not leveling, because the skew Angle will cause the dry ball temperature tube skew, and then will drop water, then just move in the machine equipment or move to the leveling road surface can be.

4, check the constant temperature and humidity test box on the right side of the water level line box is not too high, eventually leading to too much water in, if this kind of situation, people only need to adjust the water level line box. In addition, it is recommended to communicate with the manufacturer in regulating the water level line box, so that the original manufacturer can assign professional and technical personnel to the spot for maintenance, in case of abnormalities.

In the event of a problem solving problem, this is the operating principle that the operator needs to know. In the case of a clear direction, in the case of a solution, one by one according to the above reasons, in the future use of the process, but also reduce the small problems that should have occurred.